A Device for Determination of Vapour Sorption in Polymers
Abstract
The principle of the device consists in measuring the mass of vapour sorbed in a sample as a function of time using the McBain spiral balance monitored by a camera system. Absolute pressure is monitored by a capacitance pressure gauge. The maximum amount of the sample is limited to 0.5 g and the mass change sensitivity threshold is 10 μg. The device makes it possible to measure sorption kinetics with good accuracy. As an example, the kinetics of sorption of p-xylene vapours by low-density polyethylene membrane was measured and the resulting sorption isotherm was determined.Downloads
Published
2009-05-15
How to Cite
Vopička, O., Hynek, V., Friess, K., Šípek, M., & Sysel, P. (2009). A Device for Determination of Vapour Sorption in Polymers. Chemické Listy, 103(4). Retrieved from http://blog.chemicke-listy.cz/ojs3/index.php/chemicke-listy/article/view/1551
Issue
Section
Articles